Gain: 1/10,000
-
Bandwidth: 100
MHz
-
Accuracy:
0.1%
DC to 10KHz
-
Voltage
Coefficient: -0.12ppm/V
-
Temperature
Coefficient: +25ppm/oC
-
Maximum Operating Differential Input RMS Voltage:
6.3KV
-
Maximum Operating Common Mode Input RMS Voltage:
3.15KV
-
Maximum Nondestructive Differential Input Voltage: 12
KV
-
Maximum Nondestructive Common Mode Input Voltage: 6
KV
-
Output Voltage: 3.5
V peak to peak
-
Output Impedance: 50
W
-
Propagation Delay: 10
ns
-
Rise Time: 3
ns
-
Differential Input Capacitance:
0.2
pF
-
Common Mode Capacitance: 0.4
pF
-
Differential Input Resistance:
20
MW
-
Common Mode Input Resistance:
10
MW
each input to ground
-
AC Common Mode Rejection Ratio
60
Hz>100,000:1, 1 MHZ>1,000:1, 100 MHz>100:1
-
Digital
Offset Adjustment Using Momentary Up-Down Switches
|
Applications
-
Inverter And
Switching Power Supply Testing
-
Electric Car
Power System Testing
-
Compliance Testing
IEC801/1000
-
Line Monitoring
-
Computer Monitor
and TV Testing
|